LN2 Cryostat for Optical Measurements
MLN2CS-20

 
LN2 Cryostat for Optical Measurements MLN2CS-20
 
This cryostat for microscopic measurements features a synthetic quartz optical window and liquid nitrogen sample cooling. After the sample chamber is evacuated to provide vacuum insulation, the sample stage is cooled with liquid nitrogen, allowing optical measurements of the cooled sample. The distance from the outer surface of the measurement window to the sample stage is only 3 mm, enabling measurements with short-working-distance objective lenses. The synthetic quartz window supports measurements over a broad wavelength range from ultraviolet to near-infrared.

■Specifications

 Temperature Range  80 K (-193°C)–400 K (127°C) 
 Maximum Sample Size  16 mm diameter × 1 mm thickness
 Window Material / Diameter / Thickness  Synthetic quartz / 20 mm / 1 mm
 Sample Stage Size  20 mm diameter
 Distance from Sample Stage to Outer Window Surface  3 mm
 Temperature Sensor  Pt100 resistance temperature detector
 Vacuum Pressure  1 × 10−2 Pa (recommended value)